Description
Composition
100 g contains 9.1 g alkylamine, 3.6 g didecyldimethylammonium chloride, corrosion inhibitors, auxiliaries and water.
Effective Spectrum
BACTERIA TIME
Enterococcus hirae |
EN 13727:2012+A2:2015 |
2% in 5 minutes |
Enterococcus hirae |
EN 14561:2006 |
2% in 5 minutes |
Mycobacterium avium |
EN 14348:2005 |
2% in 5 minutes |
Mycobacterium avium |
EN 14563:2008 |
2% in 5 minutes |
Mycobacterium terrae |
EN 14348:2005 |
2% in 5 minutes |
Mycobacterium terrae |
EN 14563:2008 |
2% in 5 minutes |
Pseudomonas aeruginosa |
EN 13727:2012+A2:2015 |
2% in 5 minutes |
Pseudomonas aeruginosa |
EN 14561:2006 |
2% in 5 minutes |
Staphylococcus aureus |
EN 13727:2012+A2:2015 |
2% in 5 minutes |
Staphylococcus aureus |
EN 14561:2006 |
2% in 5 minutes |
Staphylococcus aureus |
EN 13727:2012+A2:2015 |
2% in 5 minutes |
Candida albicans |
EN 13624:2013 |
2% in 5 minutes |
Candida albicans |
EN 14562:2006 |
2% in 5 minutes |
Adenovirus |
EN 14476:2013+A1:2015 |
2% in 5 minutes |
Bovine viral diarrhea virus (BVDV) |
EN 14476:2013+A1:2015 |
2% in 5 minutes |
Coronavirus (FCoV) |
EN 14476:2013+A1:2015 |
2% in 5 minutes |
Herpes simplex virus (HSV-1) |
EN 14476:2013+A1:2015 |
2% in 5 minutes |
Influenza A virus (H3N2) |
EN 14476:2013+A1:2015 |
2% in 1 minute |
Norovirus (MNV) |
EN 14476:2013+A1:2015 |
2% in 10 minutes |
Poliovirus |
EN 14476:2013 |
2% in 15 minutes |
Vaccinia virus |
EN 14476:2013+A1:2015 |
2% in 5 minutes |
Chlorinated polyvinyl chloride |
Compatible |
High-density polyethylene |
Compatible |
Low-density polyethylene |
Compatible |
Polyamide |
Compatible |
Polybutylene terephthalate |
Compatible |
Polyetherimide |
Compatible |
Polyethylene terephthalate |
Compatible |
Polymethyl methacrylate |
Compatible |
Polyoxymethylene / polyformaldehyde |
Compatible |
Polyphenylene oxides / polyphenyl ether |
Compatible |
Polyphenylsulfone |
Compatible |
Polypropylene |
Compatible |
Polysiloxane |
Compatible |
Polysulfone |
Compatible |
Polyvinyl chloride |
Compatible |
Polyvinylidene fluoride / polyvinylidene difluoride |
Compatible |
Thermoplastic elastomer |
Compatible |
Thermoplastic polyurethane |
Compatible |
Ultrahigh molecular weight polyethylene |
Compatible |
Silicon carbide |
Compatible |
Aluminium |
Compatible |
Carbide |
Compatible |
Hard metal |
Compatible |
Rhodium |
Compatible |
Stainless steel |
Compatible |
Steel |
Compatible with restrictions |
Tungsten carbide |
Compatible |
Tungsten carbide (chameleon effect coating) |
Compatible |
Tungsten carbide (diamond like carbon coating) |
Compatible |
Tungsten carbide (zirconium nitride coating) |
Compatible |
Diamond (diamond like carbon coating) |
Compatible |
Diamond (gold coated) |
Compatible |
Diamond (zirconium nitride coating) |
Compatible |
Diamond grit |
Compatible |
Zirconium dioxide |
Compatible |
For your protection, wear a laboratory gown, household gloves and safety goggles. Prepare the working solution according to the dosing table. Immerse instruments completely and adhere to contact time. After disinfection, rinse instruments thoroughly under running water of at least drinking quality for 10 seconds. Visually check instruments after the disinfection for contaminants. If necessary, remove contaminants with a brush and repeat the disinfection process. Replace working solution daily or when visibly contaminated. The disinfectant solution is also suitable for use in an ultrasonic cleaning device. Adhere to reprocessing recommendations of the instrument manufacturer. A list of compatible and incompatible materials is available on our website. If you are uncertain as to the material composition of the medical device, contact the manufacturer before use.
Limitation
FOR PROFESSIONAL USE ONLY. Not for final disinfection of semi-critical and critical medical devices. Sterilize medical devices after disinfection. Do not mix with other products. For virucidal efficacy, perform disinfection on previously cleaned instruments only.
Physical Properties
pH-value : ,
Shelf Life : 24
Handling
DANGER H290 May be corrosive to metals. H302 Harmful if swallowed. H314 Causes severe skin burns and eye damage. H373 May cause damage to organs through prolonged or repeated exposure. H410 Very toxic to aquatic life with long lasting effects. P280 Wear protective gloves/protective clothing/eye protection/face protection. P301 + P312 IF SWALLOWED: Call a POISON CENTER or doctor/physician if you feel unwell. P303 + P361 + P353 IF ON SKIN (or hair): Take off immediately all contaminated clothing. Rinse skin with water/shower. P305 + P351 + P338 IF IN EYES: Rinse cautiously with water for several minutes. Remove contact lenses, if present and easy to do. Continue rinsing. P391 Collect spillage. P501 Dispose of contents/container in accordance with local and national regulations.
Intended Users
The product is intended for use by dental or healthcare professionals with specific knowledge about the disinfection of medical devices. The product is not to be used directly on patients.
Packaging
Delivery Forms :
250 ml bottle
1 litre bottle with measuring cap
2 litre bottle with measuring cap
5 litre canister
0 Review for “OROLIN® Multisept Plus | Powerful disinfectant concentrate for medical and dental instruments.”
5 Stars
0%
4 Stars
0%
3 Stars
0%
2 Stars
0%
1 Stars
0%